New features of the Mini GC Plus
The Mini GC Plus can detect more compounds than the original Mini GC. The new features that make this possible include:
The maximum temperature of the column is 160°C, offering more flexibility in designing temperature profiles.
The MEMS chip sensor can be set at either of two levels of sensitivity.
Standard sensitivity mode works well for polar compounds such as: ketones, alcohols, and esters.
High sensitivity mode works well for compounds such as halogenated alkanes and substituted aromatics, as well as mixtures with one or more compound of low concentration.
Use room air as the carrier gas
Developed by Seacoast Science, the Mini GC Plus uses an efficient, software-controlled internal pump system. This design allows our Mini GC Plus to use room air as the carrier gas.
Use Vernier’s award-winning software for column control, collection, and peak integration
With either Logger Pro or the LabQuest App, peak integration analysis and retention-time determination are built right into the software.
Classes of compounds that can be used in Vernier Mini GC Plus
COMPOUND TYPE | TYPICAL COMPOUNDS | RANGE OF ACCEPTABLE BOILING POINTS |
---|
Alcohols | C1 – C6 | 50 – 175°C |
Aldehydes | C1 – C8 | 50 – 175°C |
Amides | C3 – C4 | 150 – 250°C |
Polar-Substituted Aromatic Hydrocarbons | C6 – C12 | 100 – 150°C |
Carboxylic Acids | C1 – C4 | 100 – 150°C |
Esters | C2 – C10 | 30 – 200°C |
Ethers | C2 – C8 | 30 – 110°C |
Heterocycles | C5 – C8 | 100 – 150°C |
Halogenated hydrocarbons | C1 – C8 | 30 – 120°C |
Ketones | C3 – C8 | 50 – 175°C |
Nitriles | C2 – C5 | 50 – 120°C |
Nitroalkane/aromatic | C1 – C6 | 100 – 210°C |
Phosphonates | C3 – C9 | 180 – 210°C |
Sulfides | C4 – C8 | 90 – 150°C |
See graphs for various compounds and mixtures »
Specifications
MCCD Detector
The unique Seacoast Science MEMS Chemi-Capacitive Detector is state-of-the-art technology that allows air to be used as a carrier gas (patents pending).
The MEMS chip sensor can be set at either of two levels of sensitivity.
Standard sensitivity mode works well for polar compounds such as: ketones, alcohols, and esters.
High sensitivity mode works well for compounds such as halogenated alkanes and substituted aromatics, as well as mixtures with one or more compound of low concentration.
Carrier Gas
Use room air as the carrier gas.
Column
High-quality Restek MXT-1 Stainless (11 meter) capillary column.
Pressure Regulation
Software-controlled pressure regulation from 1–21 kPa above ambient.
Temperature Regulation
Software-controlled temperature regulation from 30–160°C
Column Temperature Ramp
Maximum of 10°C/minute.
Short Warm-up time: Fast for most labs, <10 minute max
Injection
Comes with one high-quality 1µL glass syringe, 2-7/8 inch stainless needle with non-coring beveled tip and needle guard, 1.5 inch injectable length.
Injection is direct from syringe to column
2 septa included with purchase
(~150 injections/septum)
Injection volume: 0.01 to 1.0 µL
Size
1.3 kg
10.8 cm x 13.3 cm x 19.1